All public logs

Jump to navigation Jump to search

Combined display of all available logs of DRAGON Wiki. You can narrow down the view by selecting a log type, the username (case-sensitive), or the affected page (also case-sensitive).

Logs
  • 13:42, 22 September 2023 Lennarz talk contribs created page High Voltage - ISEG power supply (Created page with "As mentioned in the previous group meeting, the new MPOD crate has been installed. We have two new modules 1x 3kV and 1x 500V. Each are 16 channels. These can accommodate both polarities. Channels 0-7 are negative and 8-15 are positive. The controller module is CC24, which can interface with EPICS. These modules can be operated via the iseg software or the web interface. The host name for the control module is dragonhv02.triumf.ca and sits behind the TRIUMF firewal...")